Pass/Fail Limit (Specification) Descriptions
6. RXA Expandor Zero Reference Level (dBV)
6. RXA Expandor Zero Reference Level (dBV)
This sets the pass/fail limits used when the expandor’s output voltage
at the 0 dB reference level is measured. Upper and lower limits must be
entered in dBV rms.
Pass/fail limits are determined by using any applicable standard, such
as:
• MSUT Specification
The test is performed and pass/fail limits are available so that the
operator can measure the RXA Expandor Zero Reference Level and
compare the result to specifications that meet his or her needs.
If the operator does not wish to compare the measurement results to
specifications, the “check” setting for pass/fail limit 15. TXA
Compressor Tracking Error >0dB (dB) can be set to “none.” See “To
Change Pass/Fail Limits” in the
Agilent Technologies 83217A Test
Software User’s Guide.
Example
If you desire the output voltage from the receiver to be −20 dBV rms ±1
dB, enter
−21 as the Lower Limit and −19 as the Upper Limit.